Warning! This is the archived 1999 Fabweb site! Here is the latest site
ECE 344 Equipment Instructions for:
The ELLIPSOMETER
Manufacturer: Gaertner
Model: L117
As described in Section 5.10.2 of Anner's Planar Processing Primer,
ellipsometers are an accurate means of measuring thicknesses and refractive indicies of
transparent films.
Quantities Used later on
- Nf - Real part of the refractive index of the film
- Ns - Real part of the refractive index of the substrate
- Kf - Imaginary part of the refractive index of the film
- Ks - Imaginary part of the refractive index of the substrate
- THICK - thickness of the film in Angstroms
- 1 Angstrom = 1E-10 meters
- P1 - First polarizer reading
- P2 - Second polarizer reading
- A1 - First analyzer reading
- A2 - Second analyzer reading
- PSI = {180deg-(A2-A1)}/2
- DEL = 360deg-(P1+P2)
- Turn on laser using the on-off key if it's not on already;
the indicator light should
come on immediately. For best results allow 15 min. for laser
warm up. The beam attenuator should be closed during warm up for
safety.
- Set the analyzer and polarizer to the desired angle of
incidence (70 deg. recommended). Both should be set to the same
angle and clamped in place.
- To change the angle of incidence: (not usually necessary.)
- Loosen the large knob (clamp screw) on the back of the arm
(behind the instrument).
- Support the arm and pull the small knob (locating pin). When
setting the polarizer arm, support it under the drum, not under
the laser.
- With the pin pulled out, move the arm to near the new setting.
Release the pin and continue moving the arm until the pin engages.
Tighten the clamp screw.
(Caution, never lower either arm to the 90 deg. setting without
first lowering the sample stage as far as it will go, since the
microspot optics will hit. To prevent this, the 90 deg. locating
pin hole is blocked. The 90 deg. setting is used only for
checking instrument alignment and for this the block may be
removed. See alignment section.)
- Place the wafer to be measured on the sample stage.
Experimentally, it has been shown that for ECE 344 purposes,
the wafers may be left in their carriers with the lid off.
The sample stage will normally be left at the proper height
for this. For more accurate measurements the wafer
should be held flat on the sample stage by vacuum and the stage
height adjusted accordingly.
- Open the beam
attenuator. Looking through the microscope, you should see the
laser spot in the center.
- Use the x-y adjustment
knobs to position the desired measurement point under the
crosshairs. The laser spot on the wafer is about 25 microns
vertically and 86, 42, or 30 microns horizontally for 70, 50, or
30 deg. angle of incidence, respectively. Turn off the
illuminator if used.
- Adjust the gain knob to give an extinction meter reading
between 150 and 200.
- Rotate the analyzer (right drum) slowly in the red numbered
region (0-90 deg.) to get a minimum reading on the meter.
- Then rotate the polarizer (left drum) slowly in the red region
(315-135 deg.) to get a lower minimum meter reading. The
numerical value of the meter reading is not important, only that
it be a minimum. However, the meter is most sensitive between 25
and 100 (above 100 it is logarithmic), so fine adjustments of the
polarizer and analyzer are easiest with the gain adjusted to give
a meter reading in the 25 to 100 range.
- Return to the analyzer (right hand) and slowly rotate the drum
to give a new minimum. This should be within the red region (0 to
90 deg.). If the new minimum is out of this range, rotate the
polarizer (left hand) to a new setting in the red region and
search for a new minimum.
- Return to the polarizer (left hand) and slowly rotate to give
a lower meter reading. (The meter gain may be adjusted to keep
the meter reading between 25 and 100.)
- Work back and forth between analyzer and polarizer to get the
lowest possible meter reading. The analyzer and polarizer
readings correspond to extinction. Adjust the gain so your minimum is
is in the 25 to 100 range as you work. The sensitivity seems highest near 100.
- Record the analyzer and polarizer settings, A1 and P1,
respectively.
- Turn the gain down to its lowest setting (counterclockwise).
- Set the polarizer to P1 + 90 deg.
- Set the analyzer to 180 - A1 deg.
- Again work back and forth between analyzer and polarizer (increasing
the gain as before) to
get the lowest possible meter readings (extinction).
- Record the polarizer and analyzer settings for the second
extinction, P2 and A2. To obtain accurate measurements, these
values should not differ by more than 4 deg. from the values
computed in 13 and 14.
Loading the Software:
The easiest way to start the software is by selecting "Ellipsometer" from
the ECE 344 customized root menu. Otherwise
- log into one of the machines in lab (like vdpauw.ece.uiuc.edu)
- if you are running remotely, you need to display on the terminal you are
sitting at.
- in an xterm of the 344 machine, type: setenv DISPLAY "name where you want
displayed."
- then in an xterm where you are sitting, type xhost +"name of 344 machine
you are logged into."
- then type the rest in the 344 machine's xterm:
- cd /usere/classes/ece344/basic
- rmb &
You will
be prompted:
"For Film program press f7/For substrate program press f8."
When you need to get from the substrate program to the general
program, or restart either program, press RESET (SHIFT-BREAK), and
then function key f3 (RUN).
Note: For ECE 344 purposes, you may go directly to the General
Program GP9. However, to be most accurate, you should make use of
this program first.
This program is used to find the refractive index values of a
substrate so that the data can be analyzed with the General
Program GP9. The General Program defaults to typical values for
silicon substrates when no values are supplied by the user. The
refractive index of silicon varies depending on quality and
doping, so for more accurate results when using silicon
substrates, the index values should be found with this program,
and used in the general program GP9.
Note: Measurements must be made on bare substrates.
DATA ENTRY: Angle of incidence F, analyzer and polarizer readings
A1, P1, A2 and P2, or PSI and DEL, where PSI = {180deg-(A2-A1)}/2,
and DEL = 360deg-(P1+P2).
DATA OUTPUT: Ns, Ks, PSI and DEL.
INSTRUCTIONS:
- Select NO PRINTING or PRINTING mode using the function keys as
prompted.
- Enter DATE and/or RETURN.
- Enter the value of F if different from the value displayed
(usually 70deg) and/or RETURN.
- Enter analyzer and polarizer readings A1 and P1 as prompted.
(If PSI and DEL values are used, press RETURN when prompted for
A1, and then enter PSI and DEL values as prompted.)
- When A1 and P1 have been entered, the program displays values
for A2 and P2 which should be within a couple degrees of your
readings. Press RETURN and then enter your values for A2 and P2
as prompted.
- The program will then prompt "A1=?" again. At this point you
can enter an additional set or sets of data and the values will be
averaged in the final computation. When you are done entering
data, just press RETURN at this prompt.
- The program now prompts "RTD=90?" This is the retardation
angle, a parameter adjustable on some ellipsometers, but not the
L117, for which it is always 90deg. Unless you are taking data on
another ellipsometer where RTD 90deg, press RETURN to continue.
This initiates the calculation, followed by a beep and the data
output. Record the values for use in the general program.
- Normally one would proceed on to the general program, but
here is how to quit the program.
Note: The ENTER key can be used in place of the RETURN key if
desired.
This program can analyze two-layer or single-layer, absorbing or
non-absorbing (transparent) films, as described earlier. Refer to
the earlier section "Modes of Operation" to select the mode you
wish to use, and determine which parameters, if any, you need to
fix. The program mode of operation is determined by parameters
that are fixed (in step 4 or 10 below), and by parameters for
which there are no values entered. For transparent layers, always
press RETURN when prompted for K. When working with single
layers, always press RETURN when prompted for the layer 2
thickness.
As the program steps through the necessary segments to initialize
the required parameters for the calculation, it identifies each
segment with a function key label (such as f1). After this
initialization process is done, function key labels are displayed
along the bottom of the screen. These are used to reenter the
initialization process as required to change the program
parameters and mode.
DATA ENTRY: Substrate refractive index Ns and Ks, top layer
refractive index Nf and Kf, layer 2 thickness, layer 2 refractive
index N2 and K2, angle of incidence F, top layer expected
thickness, and analyzer and polarizer readings A1, P1, A2 and P2,
or PSI and DEL, where PSI = {180deg-(A2-A1)}/2, and DEL = 360deg-
(P1+P2).
NOTE:
For properly made measurements, A2 ~ 180-A1, and P2 ~ P1+90. If
your readings are not within a couple degrees of the displayed
estimates, check that you entered your values properly or repeat
the measurement. Note that both A1 and P1 must always lie in the
red portion of the scales.
DATA OUTPUT: Substrate refractive index Ns and Ks, angle of
incidence F, analyzer and polarizer readings A1, P1, A2 and P2,
and/or PSI and DEL, fixed input values, expected thickness,
computed thickness and index values. For transparent films, a
listing of all possible thicknesses is optional.
INSTRUCTIONS:
- Select NO PRINTING or PRINTING mode using function keys as
prompted.
- ECE 344 students should simply press RETURN until prompted for
EXPECTED THICKNESS and skip to the next numbered step. Otherwise:
- Enter DATE and/or RETURN.
- Enter values of substrate refractive index Ns and Ks as
prompted, if different from values displayed (the program has as
default values for these Ns = 3.85 and Ks = -0.02, typical values
for silicon). The defaults are fine for ECE 344. Press RETURN after each entry.
- The program will then prompt "f1: For Fixed Top N and K press
1 before RETURN." For a single film as in ECE 344, Press RETURN.
If you are only fixing one of the values,
this will be done later under f3 (step 10). When you do not press
1 before pressing RETURN, the layer 2 thickness ("thk2") will be
fixed, and vice versa.
- Enter values for the top layer refractive index (Top N and Top
K) as prompted, if different from that displayed (the program has
a default value Top N = 1.46, typical for SiO2). Press RETURN
after each entry. (For a transparent top layer, do not enter a
value for K.)
- For two-layer films, you are then prompted with "Layer 2 Thickness=?".
Enter a value
(in Angstroms), or press RETURN for single-layer films. When you have
fixed the layer 2 thickness in step 4 by pressing RETURN without
first pressing 1, you must provide the known thickness.
Otherwise, an estimate is sufficient.
- For two-layer films, you are then prompted to enter the Layer
2 N and Layer 2 K. Press RETURN after each entry. (For a
transparent layer 2, do not enter a value for K.)
- The program then prompts "f2: PHI=70?" Enter a different
value for PHI and/or RETURN.
Thus, if you are using option (2) of the two layer transparent
film mode (i.e., fix all indices and calculate both layer
thicknesses), you must press 1 before RETURN (and press RETURN
when prompted for Kf to indicate a transparent top layer), or the
program will fix the layer 2 thickness and skip f3 (the fix Nf or
Kf option).
- The prompt "f5: Expected Thickness?" is then displayed. This
always refers to the top layer (the only layer if in the single-
layer mode). Enter the expected thickness in Angstroms.
For absorbing films, you can press RETURN with no
entry to display the thickness range choices for selection. See
the "Notes on Thickness" section for more information.
- ECE 344 students will normally go on to the next step, but if
you did not fix both Nf and Kf in step 2, and you are not
in the single-layer transparent film mode, "f3: Fix Kf or Nf:" now
appears with the instructions:
- To fix Kf: Press RETURN
- To fix Nf: Press 1 & RETURN.
Fix the desired parameter as indicated.
- The program now beeps, and is ready to accept ellipsometer
data. Enter analyzer and polarizer readings A1 and P1 as
prompted. (If PSI and DEL values are used, press RETURN when
prompted for A1, and then enter PSI and DEL values as prompted.)
To enter the program at this point later, use function key f4
(A1/P1).
- When A1 and P1 have been entered, the program displays values
for A2 and P2 which should be within a couple degrees of your
readings. Press RETURN and then enter your values for A2 and P2
as prompted.
- The program will then prompt "A1=?" again. At this point you
can enter an additional set or sets of data and the values will be
averaged in the final computation. When you are done entering
data, just press RETURN at this prompt.
- The program now prompts "RTD=90?" This is the retardation
angle, a parameter adjustable on some ellipsometers, but not the
L117, on which it is always 90deg. Unless you are taking data on
another ellipsometer where RTD 90deg, press RETURN to continue.
This initiates the calculation, followed by a beep and the data
output. You're DONE!
Note: The ENTER key can be used in place of the RETURN key if
desired.
For properly made measurements, A2 ~ 180deg-A1, and P2 ~ P1+90deg.
If your readings are not within a couple degrees of the displayed
estimates, check that you entered your values properly or repeat
the measurement. Note that both A1 and P1 must always lie in the
red portion of the scales.
After the calculation is done, the following function key labels
appear across the bottom of the screen:
f1 f2 f3 f4 f5 f6 f7 f8
Nf & Kf PHI Fix NfKf A1/P1 EXP THK SAMPLE LISTING PRT/DISP
Function keys f1 through f5 are used to enter the program at the
indicated points, as described above. Function key f6 (SAMPLE)
allows you to display and print out a sample name and/or number
(such as Wafer #15). Function key f7 (LISTING) prints out all the
possible thicknesses for several periods for transparent film
layers. Function key f8 (PRT/DISP) switches between PRINTING and
NO PRINTING modes.
- Press RESET (SHIFT-BREAK).
- Type "bye"
equipment/Ellipsometer/Instructions.html
This screen was created by Mike Fitzimmons - U of Illinois ECE Dept. -
mikef@uiuc.edu
E-mail comments and suggestions to ece344@uiuc.edu
Warning! This is the archived 1999 Fabweb site! Here is the latest site